AUDIT OF PEDIATRIC CHEST RADIOGRAPHY IN SOUTHERN NIGERIA *Ekpo E.U. Erim A.E. Egbe N.O. Chiaghanam NO, Sampson PO Department of Radiography and Radiological Science, University of Calabar, Calabar, Cross Rivers State. Email: [email protected] ABSTRACT

INTRODUCTION

Background: Technical faults represent one of the limiting factors in optimizing pediatric radiography. These faults affect the technical quality of the radiographs produced.

Audit of technical faults is important in radiography as they form part of the assessment criteria for establishing quality of patient care received in Radiography department. Audits are particularly important in pediatric radiography because of the 1 challenges encountered with pediatric examination .

Objectives: We sought to audit the common technical faults in pediatric chest radiography in South-South Nigeria towards identifying and adopting remedial measures that would ensure improved Radiography service delivery in the region. Methods: A total of 621 pediatric (<12years) chest radiographs were independently and subjectively assessed for the presence or absence of faults by two Quality Assurance (QA) radiographers with 12±2 years average working experience. Parameters assessed were rotation, collimation, blurring, exposure factors, artifacts and identification. Result: Findings show that out of 621 pediatric radiographs assessed, 551 had one or more faults. Out of 551 radiographs with faults, 450 (81.7%) were poorly collimated, 159 (28.3%) were either over or underexposed and 71 (12.9%) had artifacts. Conclusion: Patient rotation, blurring, presence of artifacts, wrong identification and choice of exposure factors are the common faults in pediatric chest radiography in the area studied. Training and retraining of radiographers, installation of dedicated pediatric facilities, legislation monitoring and evaluation of practice and regular QA will help in optimizing pediatric radiography in South-South Nigeria.

Pediatric patients in this case refer to children below 12years of age. Individual between the 0 - <12 year have limited or no awareness of the changes in their environment. Their level of cognition varies and directly influences their actions and reaction to change around them, determining also the degree of cooperation obtainable from them during radiological investigations1. Technical faults in pediatric radiography may be due to the peculiar presentation of these children and their inability to comply with radigraphers’ instruction for the production of diagnostic quality radiographs. The significance of these faults vary from minimal to severe. The resultant effect may generate bad publicity, losses to the hospital, wrong and delayed diagnosis, and increased radiation risk arising from 2,3,4 ensuring repeat of radiographic examinations . The potential of ionizing radiation to cause damage to biological tissues requires that its use be justified and then optmized4,3,5. Justification implies that the benefit from any radiographic procedure must outweigh the risk. One way of reducing the exposure to risk is by optimizing the procedures involved optimization is aimed at producing diagnostic quality images at 6 minimum radiation dose and cost .

Keywords: Technical faults, pediatrics, radiographer, collimation, exposure, artifacts, Identification, rotation. Nigerian Journal of Medical Imaging and Radiation Therapy

6

Vol 2 No 1 November 2013

Audit of Pediatric Chest Radiography in Southern Nigeria

This is particularly important in pediatric radiography because children are more susceptible to the effects of ionizing radiation owing to their highly primitive, rapidly dividing and proliferation cell content, careful and correct radiographic examination reduce repeats, waiting time, radiation does, risk and the cost of examination7,8. Plain chest x-ray is one of the common x-ray examinations for evaluation of the pediatric chest. This may be due to the ease, availability of the equipment and low cost of the examination. However, the examination challenges the technique and ability of the radiographer and places an added burden on the radiographer because children are usually uncooperative and may not follow radiographer’s instructions during examination. This is further compounded by the fact that a child’s reaction to illness is dependent on their mental, physical, social, cultural and environmental factors which in turn, is influenced by their level of cognition1. Radiographers, therefore, need to awaken their creativity and introduce new ways of coping with any difficulty they may encounter during the course of the examination. METHODS An assessment of pediatric (0-12years) chest radiographs was carried out by two quality assurance radiographers of 12 ± 2year’s average working experience. 691 radiographs were retrieved from the archives of Radiology departments of teaching hospitals in South-South geopolitical zone of Nigeria. A total of 621 pediatric chest radiographs were independently and subjectively assessed for the presence or absence of common technical faults. The conditions of viewing were optimized for the assessors with the use of viewing bow with 2 luminance values (1500cdm ). Room ambient lighting was in line with international recommendations.

Nigerian Journal of Medical Imaging and Radiation Therapy

Parameters assessed were patient rotation, beam collimation, motion and respiratory blur, exposure factors, film identification and artifacts. Collimation was assessed based on the European Guidelines on Quality Criteria for Diagnostic Radiographic Images in Pediatric 14 chest radiography . A meter rule was use to measure the distance between the lateral chest wall soft tissue and area of the film exposed. Also, measurements were made from the costophrenic angles and the distal portion of the film exposed and from the first rib to the proximal portion of the exposed radiograph. Radiographs with area of exposure of 2cm from area of interest were regarded as poorly collimated. Rotation was assessed by the distance of both clavicles from the midline. Radiographs with both clavicles not equidistant from the midline were regarded as rotated. Exposure factors were assessed by the assessed by the presence of unsharpness on the radiograph. Radiographs without any or all of patient name, number and anatomical marker were regarded as not properly identified. Any opacity, mark or appearance on the radiograph not representative of the patient’s area of interest was classified as an artifact. Parameters were either given a score of 0 for presence of fault or 1 for absence of such fault. Rule of majority (radiographs with the same opinion by the two radiographers) was used to confirm the presence of the technical faults on the radiographs. Film for which the assessors did not agree on the presence or absence of any fault were not included. A total of 621 films assessed met the criterion leaving 70 for which assessors disagreed. RESULTS Table 1: Technical Faults and Scores Technical Faults Score obtained/Number of films obtaining score: 01 Mot ion/Respiratory blur: 159392 Patient rotation: 142409 Over/underexposure: 123428 Identification (name, number and marker): 141410 Collimation: 450101 Artifact: 71 480

7

Vol 2 No 1 November 2013

Audit of Pediatric Chest Radiography in Southern Nigeria

Table 1: Distribution of and Scores compared with the mean ESD of this work.

Figure 1: Percentage distribution of faults in the dataset

Results show that out 621 pediatric radiographs assessed, 551 had faults out of this, 450 (81.7%) were poorly collimated, 159 (28.9%) were blurred, 142 (25.8%) were rotated, 141 (25.6%) were not properly identified, 123 (22.3%) were either over or underexposed and 71 (12.9%) had artifacts.

Nigerian Journal of Medical Imaging and Radiation Therapy

8

Vol 2 No 1 November 2013

Audit of Pediatric Chest Radiography in Southern Nigeria

DISCUSSION: Good radio diagnosis begins with production of good quality radiographs having sufficient contrast, resolution, low level of unsharpness and devoid of artifacts. The complexity and varying level of pediatric coperation especially in poor health conditions limits the ability of radiographers to achieve this. Although radiographic technique may not vary directly from that used for adults, modification of radiographic technique by the radiographer is often necessary to overcome the demands of varying child size, conditions and types of pediatric patients. It is often difficult for radiographers with limited experience to modify technique to suit every child at a level appropriate to the child and this is compounded by the fact that in stressful situations children will often be highly uncoperative. These tend to add to the pressure to obtain good quality films, often leading to increased incidence of faults. These faults have consequences ranging from repeat procedure to increases in lost time, radiation dose to workers and wrong diagnosis 2,3,4. amongst others. Pediatric chest radiography therefore presents challenges to radiographers trying to make a compromise between image quality and radiation dose, and between examination time and role ambiguity.

Poor collimation (field size 2cm than the area of interest) accounted for 81.7% of the total faults (Figure 1). This may be expected as radiographers would prefer using a large radiation field to avoid cutting off anatomical details which may necessitate repeat, increasing radiation doses to patients, cost of examination and examination time. This fell short of CEC guidelines which recommended the maximum field size tolerance to be less than 2cm greater than the area of interest and to be further reduced to a 14 tolerance of 1cm in neonates . Although collimation is an appropriate way of reducing radiation dose to the patient, its application requires the radiographers to apply precise knowledge of external anatomical landmarks to the pediatric patient being examined, good patient immobilization and nature of the 10 1 underlying disease. Corks et al , Hardy and Boynes have identified poor collimation as a recurring phenomenon in pediatric radiography. Patient rotation and motion blur accounted for 25% and 28.9% of faults respectively (Figure 1). Both faults in some cases occurred in isolation or concurrently. They were probably due to wrong positioning, movement of patient and non-control of respiration. These affect not only the technical quality of the radiograph but also cause misrepresentation as well as distortion in the final image which may make interpretation difficult. 15

However, no matter the condition of the child, justification and optimization of practice are paramount in their radiographic examinations to reduce the potential effect of radiation and to reduce these faults. Our study sought to assess the common technical faults on plain pediatric chest radiographs in SouthSouth Nigeria using a representative sample of pediatric chest radiographs. The results show poor collimation, motion blur, patient rotation, improper or lack of film identification, wrong choice of exposure factors and artifacts to be common technical faults in pediatric chest radiography in South-South Nigeria (Table 1). 3,4

Nigerian Journal of Medical Imaging and Radiation Therapy

The CEC , Cooks et al, observed that incorrect radiographic positioning and inadequate mobilization of pediatric patients were the most frequent causes of inadequate and poor quality pediatric images. Correct positioning of patient of patient in supine position prior to exposure and use of restraints in difficult circumstances has proven to be beneficial in reducing these faults13. The selection of exposure factors has also been a challenge. This is attributable to the varying sizes of pediatric patients and different clinical conditions. It is not made any easier by the constraint of imaging with facilities that do not have automatic exposure control (AEC). The image quality and radiation dose to patients are determined by the exposure factors used and image receptor detective quantum efficiency. About 22.3% of the radiographs were either over or under exposed (Figure1). This might have been due to wrong choice of exposure factors.

9

Vol 2 No 1 November 2013

About 25.6% of the radiographs had no proper identification (name, number and anatomical marker). The urgency with which pediatric examination are performed may be implicated in this. Wrongly placed anatomical markers and obscuring of anatomical details by markers and opaque legends may cause wrong diagnosis or necessitate repeat which increases dose to patients, patient waiting time, cost of examination and thus stresses the patient and radiographer.

Also, attitudinal change particularly towards pediatric patients and their examinations by the installation of dedicated pediatric facilities, legislation, monitoring, and evaluation of practice by the regulatory authorities will help in the optimization of pediatric radiography and reduce these faults to a minimum in South -South Nigeria. CONCLUSION

Artifacts are not uncommon findings in pediatric radiography and result from poor patient and room preparation. Although it is the responsibility of the radiographer to ensure all radio-opaque materials are removed from the region of interest, some of them may elude the radiographer and appear on the processed radiograph. Others may come from the darkroom. Presence of artifacts is the least common fault (12.9%) in pediatric radiography as observed in our study (figure 1).

The most common technical faults in pediatric radiography in the area studied are poor bean collimation, motion blur, and patient rotation, wrong choice of exposure factors, poor identification and presence of artifacts. These faults result from peculiarity of children especially in ill situations and other factors related to image productions. QA programs in the system, individual radiation workers adherence to guidelines and use of dedicated pediatric radiography units as well as modalities that will increase the child’s level cooperation may be beneficial in eliminating these faults.

Errors such as patient rotation, poor collimation, and presence of artifacts, improper identification, and wrong choice of exposure parameters, movement and respiratory blurring among others have been documented in literature 1,9,10.

REFERENCES

Several approaches have been employed to combat these errors including immobilization or restraints, short exposure time, optimal filtration, collimation and use of fast film-screen combination though not 12,13,16 respiration is either absent or unsatisfactory . Proper immobilization of the child without restraint may be dependent on the trust the child places on such a radiographer. Pediatric radiography doses have been reported, but studies of this nature have not been 17 carried out in Nigeria . These necessitate the need to investigate the common technical errors in pediatric chest radiography in South-South Nigeria in order to implement appropriate corrective measures. The results outlined above are evidence of need for QA action and regulation of practice in the system. Several studies Egbe et al19, Ekpo et al20 made recommendations for introduction of viable QA in the past in the face of similar technical findings even in adult radiographs and poor equipment maintenance policies. These recommendations are upheld in the current study. Nigerian Journal of Medical Imaging and Radiation Therapy

1. Hardy S and Boynes S. Paediatric Radiography. Blackwell publishing, 2003 pp 16-20. 2. Gully HR (2001). Diagnostic Error in Accident and Emergency Departments. Injury 15:403-269. 3.

International Atomic Energy Agency (IAEA). Radiation Protection In Pediatric Radiography: Radiation Protection of Children in Screen Film Radiography, LO3, Available at: HYPERLINK "http://rpop.iaea.org/RpoP/RPoP/content/ndex. h t m " h t t p : / / r p o p . i a e a . o rg / R p o P / R P o P / content/index.htm. Assessed: December 2012

4. Thompson MA, Hattaway MP, Hall JD, Dowd SB. Principle of Imaging Science and Protection. W.B. Saunders Company. 1994; Pp. 222-224, 430. 5. Alzen G and Benz-Bohn G. Radiation protection inpediatric radiology. Dtsch Arztebl, 2011; 108 (24): 407-414.

10

Vol 2 No 1 November 2013

Audit of Pediatric Chest Radiography in Southern Nigeria

6. Seibert J A. Trade offs between image quality and dose. Pediatric Radiology 2004; 34 (SUPPL. 3): S183-S195.

16. Mohammadain KEM, Azevedo ACP, Radiation dose survey in conventional pediatric radiology. J.Sc. Tech, 2009;10(2): 175-184.

7. Whitley AS, Sloane C, Hoadley G, Moore AD, Alsop CW. Clark’s positioning in Radiography. 2nd edn. Arnold publisher London. 2005. Pp. 4378, 448 -59.

1 7 . U g w u A C , E r o n d u O F, U m e a n o , U B , Psychosocial stress and its predictors, among radiographers in South Eastern Nigeria. The South African Radiographer, 2011; 49(2): 11-14.

8. United nation Scientific Committee on effect of atoni radiation (UNSEAR). Report (2000): Sources and effect of ionizing radiation. Vol. 1 Assessed: December 2012.

18. Egbe NO, Inah GB, Azogor WE, Chiaghanam N.O, Ikamaise VC. Good radiographic practice: The Nigerian experience and the CEC recommendations European Journal of Radiography (2009) 1,147e150.

9. Bontrager KL and Lampignano JP (2010) Textbook of Radiographic Positioning and Related Anatomy. 7th edition. Mosby pp. 10. Corks JV (2001). Key factor in the optimization of pediatric x-ray practice. British Journal of Radiology, 74:1032-40.

19. Ekop EU, Egbe NO, Azogor WE, Inyang SO, Upeh ER. Challenges of radiological equipment management policies in some Northern Nigerian Hospitals. The South African Radiographer, 2013; 51(1):19-22.

11. Bontrager KL and Lampignano JP (2010). Radiographic Positioning and Related Anatomy: Workbook and Laboratory Manual. Mosby pp. 12. Cowling C. The role of the radiographer in dose reduction for pediatrics ISSRT publication. Available at: HYPERLINK " http://www.imagegently.org" www.imagegently.org. Assessed December 20, 2012. 13. Shribm S and Denton E. Delivering Quality Imaging Services for Children: a report from the National Imaging Board. Gateway reference number: 13732. 14. Cook, JV, Pettet A, Shah, K. et al. (1998) Guidelines on Best Practice in the X-ray imaging of children: A manual For all X-ray Department Queen Mary’s Hospital for Children, the St. Helier NHS Trust, Carshalton, Surrey and The Radiological Protection Centre, St. George’s Healthcare NHS Trust, London. 15. European Guidelines on Quality Criteria for Diagnostic Radiographic Images in Paediatrics, J u l y 1 9 9 6 . E U R 1 6 2 6 1 . Av a i l a b l e a t : http://www.cordis.lu/ fp5Nigerian Journal of Medical Imaging and Radiation Therapy

11

Vol 2 No 1 November 2013

Audit of Pediatric Chest Radiography in Southern Nigeria.pdf ...

There was a problem previewing this document. Retrying... Download. Connect more apps... Try one of the apps below to open or edit this item. Audit of ...

3MB Sizes 2 Downloads 134 Views

Recommend Documents

The Role of Tracheotomy in Weaning - CHEST Journal
Care Medicine, Allergy, and Clinical Immunology, 96 Jonathan. Lucas St ... imposed on the respiratory system and thereby promote weaning from ... study designs. Most studies are retrospective, although a few prospective studies have been performed. M

A randomised crossover trial of chest physiotherapy in ...
Jun 18, 2009 - Email: [email protected]. Phone: ++ 44 131 242 1921. Fax: ++ 44 131 242 1870. Short Title: Physiotherapy in bronchiectasis.

in southern jordan
Jebel Rabigh appears to be the most specialised (80% of the tools are perforators). In general, the lithic industries from these three southern Jordanian sites are ...

in southern jordan
used for the beads was amazonite, the blue-green variety of the mineral microcline. .... assemblages the borers are the most abundant tools, Amuq 2 points are more ... before or after being drilled on both sides producing a bi-. PROCESS 1.

DOI: 10.1378/chest.123.4.1266 2003;123;1266-1275 Chest Cheryl L ...
should be relegated to the place of high-tidal volume ventilation and liberal transfusion practices. The Best Evidence: Low-Dose Dopamine. Administration in Oliguric Patients With. Sepsis Syndrome Is Not Therapeutic. Despite the widespread use of low

Oscillatory chest compression device
Jan 14, 2002 - (63) Continuation of application No. 08/661,931 ... See application ?le for complete search history. Primary ..... Product Brochure, “Percussionaire® Corporation Presents .... Generator 3 may be con?gured as a mobile unit.

Oscillatory chest compression device
Jan 14, 2002 - N. Gavriely et al., “Gas Exchange During Combined High and LoW Frequency Tidal Volume Ventilation in Dogs,” in. Progress in Respiration ...

Oscillatory chest compression device
Jan 14, 2002 - (Commued). 602/ 13. See application ?le for complete search history. ..... F. Ohnsorg, “A Cost Analysis of HighiFrequency Chesti ..... mucous and other secretions to build up in a person's lungs. .... mobile unit shoWn in FIG.

SOUTHERN DISTRICT OF NEW YORK
YouTube is a site Where users are able to upload and broadcast videos .... 10. Nonprofit organizations like the American Red Cross, the WWF. (formerly the ..... social network, where people from around the world can easily share video and .-.

Head of Internal Audit - IITA
May 10, 2016 - IITA is one of the world's leading research partners in finding solutions for hunger, ... expanding Business Incubation Platform (BIP). ... Monitor implementation by management of all recommendations made by the Internal Audit ... The

Interpretation of the Chest...
Jul 23, 2007 - ... important method of chest imaging, providing an easily accessible, cheap, ..... and tumour (especially bronchioloalveolar cell carcinoma).

Phonology in Syntax Tommi Leung University of Southern California ...
Aug 8, 2007 - Kwangwoon University, Seoul ... University of Southern California .... (v) Syntactic chains as defined by the list of occurrences in phrase ...

DOI: 10.1378/chest.123.4.1266 2003;123;1266-1275 Chest Cheryl L ...
located on the World Wide Web at: The online version of this article, along with updated information and services, is. ISSN: 0012-3692. ... history and physiology of low-dose dopamine administration and discuss the reasons why dopamine is not clinica

Request for Proposal to undertake Audit of MGNREGA in Tripura.pdf ...
Request for Proposal to undertake Audit of MGNREGA in Tripura.pdf. Request for Proposal to undertake Audit of MGNREGA in Tripura.pdf. Open. Extract.

An Audit of the Processes Involved in Identifying ... - Semantic Scholar
The Commission for Racial Equality (Special Educational Needs. Assessment in Strathclyde: Report of a Formal Investigation,. CRE, London, 1996) highlighted ...

Empanelment for Perpetual Audit of Stores Inventory in MCL (2017 ...
Empanelment for Perpetual Audit of Stores Inventory in MCL (2017-18).pdf. Empanelment for Perpetual Audit of Stores Inventory in MCL (2017-18).pdf. Open.

SOUTHERN DISTRICT OF NEW YORK
California at Berkeley who has been teaching for over 40 years. Her lectures have ... In addition to posting lectures, many colleges and universities are also.

Forty years of silvicultural management in southern ...
>30 cm) lying abandoned on the forest floor were quantified ... Location, harvesting year and area of the sampled management units (MU). Management unit.

pollination relationships in southern spanish ...
This paper investigates pollination relationships in a Mediterranean scrub ... Data on the reproductive biology of the species which are not included in this paper .... A statistical program in the BMDP series (Dixon 1981) entitled P2M was used ...

A review of marine phylogeography in southern Africa
Mar 16, 2011 - southern African marine realm is one of the world's most interesting environments in which to study the ..... occurred at least 16 million years ago,4 whereas congruent regional ... Science Association, the Agulhas and Somali Current L