REVIEW OF SCIENTIFIC INSTRUMENTS

VOLUME 71, NUMBER 10

OCTOBER 2000

A modified dual-slope method for heat capacity measurements of condensable gases S. Pilla,a) J. A. Hamida, and N. S. Sullivan Department of Physics, University of Florida, Gainesville, Florida 32611

共Received 13 March 2000; accepted for publication 27 June 2000兲 A high resolution nonadiabatic method for measuring the heat capacity (C P ) of bulk samples of condensable gases in the range of 7.5–70 K is described. In this method C P is evaluated by directly comparing the heating and cooling rates of the sample temperature for two algebraically independent heat pulse sequences without explicit use of the thermal conductance between sample and thermal bath. A fully automated calorimeter for rapid measurement of C P of molecular solids utilizing this technique is presented. The technique along with the automated calorimeter with a provision to apply external electric and magnetic fields is particularly useful for the study of continuous phase transitions in molecular solids as well as field induced changes in the heat capacity. © 2000 American Institute of Physics. 关S0034-6748共00兲02710-6兴

eter兲 must be short compared to ␶ 1 . It is this latter constraint which restricts the applicability of this method to small samples with sufficiently large thermal conductivity such as pure metals below 20 K. A slightly less sensitive method involving a fixed heat input followed by a temperature decay measurement is the relaxation method.7,14,15 In this method the sample is raised to an equilibrium temperature above the thermal reservoir and then allowed to relax to the reservoir temperature without heat input. A recent improvement of this technique utilizing advanced numerical methods was provided by Hwang et al.12 The relaxation method requires extensive calibration of the heat losses of the sample as a function of temperature difference between the reservoir and the final sample temperature and relies on accurate, smooth temperature calibration of thermometers. The technique depends heavily on the numerical techniques used to determine many equilibrium heat losses during the heating portion of the relaxation cycle to obtain good temperature resolution of the heat capacity changes. Though Hwang et al.12 addressed the problems arising from large ␶ 2 , this method still fails to provide the accurate values of C P of large samples with poor thermal conductivity at temperatures above ⬃20 K. A variation of the relaxation method called dual-slope method, was first discussed by Riegel and Weber.8 In this method an extremely weak thermal link to the reservoir is used and the temperature of the sample is recorded over a 10 h cycle while heating at constant power for one half of the cycle, and then allowing the sample to relax with zero heat input. The heat loss to the reservoir and surroundings can be eliminated from the calculation of C P (T) using this technique provided the sample, sample holder, and the thermometer are always in equilibrium with each other 共the reason for long 10 h cycle to cover 3 K range兲, and the reservoir temperature is held constant over the 10 h cycle. A further modification of this technique is a hybrid between the ac method and the dual-slope method,9 which reduces the duration of the cycle to about 2 h. Although the dual-slope method is

I. INTRODUCTION

The adiabatic heat-pulse calorimetry has been widely used to investigate the thermodynamic properties of materials for more than a century. The applied principle, C ⫽䉭Q/䉭T, which describes the heat capacity C of the sample, is determined by the pulse heat 䉭Q supplied to the sample under adiabatic conditions and the temperature rise 䉭T, is well known.1,2 Due to the inherent simplicity as well as the general applicability independent of the sample thermal conductivity, this method is the most favored choice for heat capacity measurements of condensable gases which have poor thermal conductivity in their low temperature solid phase.3–5 Although the traditional adiabatic calorimetry has high precision and can be used to determine the latent heat at strong first-order transitions, it is very difficult to achieve the resolution needed to characterize the temperature dependence of C P (T) close to the critical temperature T c for a second-order transition. Also, because of the inherent limitations on achieving the ideal adiabatic conditions at low temperatures and the long time required to cover a few tens of kelvin temperature range with a reasonable number of data points, new user friendly nonadiabatic techniques6–13 with excellent sensitivity are needed to study the heat capacity of condensable gases. Among these new techniques, the most sensitive method is the ac method devised by Sullivan and Seidel.6 While the ac method allows one to obtain accurate values of C, it suffers from the fact that it normally must be used for small samples at low ac frequencies and over limited temperature ranges 共typically T⬍20 K兲. In order to keep the sample in equilibrium with the heater and thermometer during one cycle, the time constant for thermal relaxation ( ␶ 1 ) of the sample plus sample holder 共or calorimeter兲 to the thermal reservoir must be short compared to the period of the driving flux. In addition, the samples’ internal equilibrium time constant ( ␶ 2 ) with the sample holder 共or calorima兲

Electronic mail: [email protected]

0034-6748/2000/71(10)/3841/5/$17.00

3841

© 2000 American Institute of Physics

Downloaded 19 Dec 2001 to 132.239.1.232. Redistribution subject to AIP license or copyright, see http://ojps.aip.org/rsio/rsicr.jsp

3842

Rev. Sci. Instrum., Vol. 71, No. 10, October 2000

Pilla, Hamida, and Sullivan

very elegant and easy to implement, the technique has only been implemented for small samples 共typically less than 0.5 g兲 with good thermal conductivity and at temperatures less than ⬃20 K. The success of this method depends heavily on achieving very good thermal equilibrium between the sample, sample holder, and the thermometer, and for large samples with poor thermal conductivity 共i.e., large ␶ 2 ) this method may also fail when we consider only the first order approximation of the heat balance equations of Riegel and Weber.8,9,12 In the case of condensable gases, the sample size must be at least a few grams to reduce the spurious effect of sample condensed in the fill line and the heat leak through this tube. Due to the low thermal conductivity of these samples, in particular for powdered samples which may not wet the calorimeter walls, the ␶ 2 can be very large leading to the failure of most of the above techniques. The present article describes a modified dual-slope method in which C P is evaluated by directly comparing the heating and cooling rates of the sample temperature for two algebraically independent heat pulse sequences without explicit use of the thermal conductance between sample and thermal bath. Although in principle we measure the heat capacity C S of the condensed sample along its own saturated vapor pressure curve, in general for solids with face centered cubic or hexagonal lattice 共H2 , N2 , CO, etc.兲 the difference between C S and C V does not exceed the measurement uncertainties for temperatures below 30 K.16 For most of the condensable gases, except very near the melting temperature, the correction for the saturated vapor pressure is negligible and C S is usually assumed to coincide with C P at P⫽0. 17 Also, the correction C P ⫺C V (⬍5%) may be neglected for T⬍25 K in the case of N2 , CO, and N2 –Ar mixtures, and for T⬍8 K in the case of H2 . For the specific geometry of the calorimeter that we used, which is most suitable for the heat capacity measurements of condensable gases in the presence of external electric or magnetic fields, higher order heat balance equations are calculated, which can easily be adapted for other sample configurations as well. Because of the explicit consideration of the higher order equations, the problem of large ␶ 2 is naturally addressed and during the heating and cooling cycles one does not require the sample to be in equilibrium with the thermometer, or in other words, the thermometer need not necessarily record the actual sample temperature. Due to this freedom we can obtain C P of samples with poor thermal conductivity very rapidly. II. THEORY

To model the thermal response of a heat-pulse calorimeter 共Fig. 2兲 for heat capacity measurements, with the schematic diagram shown in Fig. 1, where the calorimeter is heated with the application of heater power P(t) to some desired temperature T max and then allowed to cool down, we start with the following set of heat balance equations P 共 t 兲 ⫽C ⬘ T˙ h⬘ ⫹␭ s 共 T h⬘ ⫺T h 兲 ⫹␭ r 共 T h⬘ ⫺T 0 兲 and

共1兲

FIG. 1. Simplified diagram of the thermal system of sample, sample holder, and thermal bath. The thermal link between the outer conductor plus heater of heat capacity C ⬘ at temperature T ⬘ and temperature-controlled reservoir bath at T 0 is represented by the thermal conductance ␭ r . ␭ r may be controlled by changing the He exchange gas pressure. The heater input power P(t) is obtained from a wave form generator and a suitable power amplifier. The thermal link 共through the sample兲 between the inner conductor plus thermometer well of heat capacity C at temperature T and the outer conductor is represented by the thermal conductance ␭ s .

0⫽CT˙ h ⫹␭ s 共 T h ⫺T h⬘ 兲 ⫺ P 0 共 T h 兲

共2兲

for heating, and 0⫽C ⬘ T˙ ⬘c ⫹␭ s 共 T c⬘ ⫺T c 兲 ⫹␭ r 共 T ⬘c ⫺T 0 兲

共3兲

0⫽CT˙ c ⫹␭ s 共 T c ⫺T c⬘ 兲 ⫺ P 0 共 T c 兲

共4兲

and

for cooling, where C and T are the heat capacity and the temperature of the thermometer well plus the inner conductor respectively, C ⬘ and T ⬘ are those of the outer conductor plus the heater coil. From these equations we need to obtain the quantity (C⫹C ⬘ ) which is the net heat capacity of the calorimeter. Once the sample is condensed in the calorimeter, both C and C ⬘ will be modified but the above equations are still valid. P 0 (T) is the parasitic stray heat due to the conduction of heat through the fill line, the stainless steel suspension rod, and the thermometer leads as well as radiation from the pumping line. From the geometry of the calorimeter 共Fig. 2兲 it is reasonable to assume that this stray heat affects only the inner conductor heat balance equations 关Eqs. 共2兲 and 共4兲兴. Since the thermometer is firmly coupled to the inner conductor through the thermometer well, the temperature T is what the thermometer practically records. Hence, we need to eliminate T ⬘ to obtain (C⫹C ⬘ ). For T h ⫽T c Eqs. 共1兲 and 共3兲 reduce to P 共 t 兲 ⫽C ⬘ 共 T˙ h⬘ ⫺T˙ ⬘c 兲 ⫹ 共 ␭ s ⫹␭ r 兲共 T ⬘h ⫺T ⬘c 兲 .

共5兲

From Eqs. 共2兲 and 共4兲 and their first derivatives we obtain 共 T h⬘ ⫺T c⬘ 兲 ⫽

C 共 T˙ ⫺T˙ c 兲 , ␭s h

共 T˙ ⬘h ⫺T˙ ⬘c 兲 ⫽ 共 T˙ h ⫺T˙ c 兲 ⫹

C 共 T¨ ⫺T¨ c 兲 . ␭s h

共6兲 共7兲

Downloaded 19 Dec 2001 to 132.239.1.232. Redistribution subject to AIP license or copyright, see http://ojps.aip.org/rsio/rsicr.jsp

Rev. Sci. Instrum., Vol. 71, No. 10, October 2000

Dual slope heat capacity

3843

After eliminating T h⬘ , T c⬘ and their time derivatives from Eqs. 共5兲, 共6兲, and 共7兲 we obtain P 共 t 兲 ⫽ 共 C⫹C ⬘ 兲共 T˙ h ⫺T˙ c 兲 ⫹ ⫹

C␭ r 共 T˙ h ⫺T˙ c 兲 . ␭s

CC ⬘ 共 T¨ h ⫺T¨ c 兲 ␭s 共8兲

Clearly, Eq. 共8兲 reduces to the first order derivations of Riegel and Weber8 for ␭ s Ⰷ1. However, for condensed gases, in particular for powdered samples, this condition is never met. In the present geometry, for a sufficiently weak thermal link ␭ r where ␭ s Ⰷ␭ r , one can ignore the last term in Eq. 共8兲. To solve the remaining equation exactly for C⫹C ⬘ , we need to obtain Eq. 共8兲 for two algebraically independent pulse sequences of P(t). III. EXPERIMENT

Figure 2 shows the design of the calorimeter optimized for the above technique. The thermal reservoir is made up of a brass vacuum can 共18 cm long, 4 cm diameter兲 with a manganin wire wound uniformly on the entire length of the outer surface. A dilute solution of GE varnish is used to soak the cloth insulation of the manganin heater wire to provide good thermal contact with the vacuum can. A grooved copper tube 共4 cm long兲 is soldered to the top flange, which can be used for heat sinking the heater and thermometer wires before connecting them to the calorimeter. A 122 cm long stainless steel tube 共0.96 cm diameter兲 is soldered to the top flange to pump the vacuum space as well as support the vacuum can. Copper radiation baffles are placed inside this tube at regular intervals to reduce the thermal radiation. In addition, the top flange supports up to ten individual vacuum feedthroughs 共not shown兲, which are thermally anchored to the flange through a STYCAST 2850FT vacuum seal which has good thermal conductivity at low temperatures. The calorimeter consists of two concentric, thin walled, gold plated, Oxygen-free high-conductivity copper cylinders 10.2 cm in length providing a 1 mm cylindrical shell space for condensing the samples. This shell space is vacuum sealed at both ends with homemade 1 mm thick STYCAST 2850FT O rings.18 STYCAST 2850FT is particularly suitable because of its smooth change in heat capacity and thermal expansivity with temperature.19 The outer conductor 共2.7 cm inner diameter and 0.2 mm wall thickness兲 with a brass heater wire uniformly wound on the entire length of its outer surface serves as a radial heating element. Uniform winding of the heater wire is very crucial in eliminating longitudinal heat flow thereby reducing the temperature gradient along the length of the calorimeter. For good thermal contact with the outer cylinder, a dilute solution of GE varnish was applied to the heater wire. In the middle of the inner conductor 共2.5 cm diameter, 0.4 mm wall thickness兲 a thin copper ring as well as a copper thermometer well are soldered 共see Fig. 2兲. The larger wall thickness 共0.4 mm instead of 0.2 mm兲 for the inner cylinder is necessary to withstand the pressures generated when the solid samples melt at high temperatures. The copper ring sandwiched between two threaded nylon wings

FIG. 2. Cross section of the thermal reservoir and calorimeter. 共a兲 Top flange with vacuum feedthroughs, 共b兲 brass vacuum can with highly uniform heater wire wound on the outer surface, 共c兲 1.6 cm diameter copper tube with grooves on the outer surface, 共d兲 STYCAST 2850FT seal 共Ref. 18兲 共e兲 outer conductor with highly uniform heater wire wound on the outer surface, 共f兲inner conductor, 共g兲 copper thermometer wells soldered to the walls, 共h兲 threaded stainless steel rod, 共i兲 0.5 mm thick copper ring soldered to the inner conductor, 共j兲 threaded nylon wings, 共k兲 sample fill line, 共l兲 copper radiation baffles, and 共m兲 stainless steel tube for support as well as pumping the vacuum space.

supports the calorimeter when suspended from the top flange with the help of a threaded stainless steel rod. This arrangement not only reduces the unwanted heat leaks to the calorimeter but also provides excellent mechanical stability when the free end of the stainless steel rod is slipped into the post at the bottom of the vacuum can 共see Fig. 2兲. The 1.6 mm diameter cupronickel alloy tube serving as the sample fill line is coiled into a three turn spring 共not shown兲 before soldering it to the inner cylinder. To be able to apply an electric field across the sample cell when necessary, it is important to ground the outer conductor 共to reduce electrical interference from the heater current兲 and apply the high voltage to the inner conductor. Hence, the coiled section of the fill line is isolated electrically as well as thermally from the remaining length with the help of a vacuum joint made up of a wider cupronickel tube, cigarette paper, and STYCAST seal. To assist sample condensation, the manganin heater wire is wrapped on the fill line tube outside of the vacuum can. Two factory calibrated germanium resistance thermometers20 placed inside the thermometer wells are used

Downloaded 19 Dec 2001 to 132.239.1.232. Redistribution subject to AIP license or copyright, see http://ojps.aip.org/rsio/rsicr.jsp

3844

Rev. Sci. Instrum., Vol. 71, No. 10, October 2000

Pilla, Hamida, and Sullivan

FIG. 4. Heat capacity of the calorimeter filled with pure N2 sample. Data are obtained for two pulse sequences with respective T 0 values separated by 2 K.

FIG. 3. Traces of T(t) and T 0 (t) for two heat pulses when the calorimeter is filled with pure N2 sample. Useful interval for obtaining heat capacity is shown between two dashed lines. The difference between the trace of T 0 and the baseline of T is due to the parasitic stray heat P 0 .

for recording the temperatures of the calorimeter and the thermal reservoir. The entire assembly is placed in a second vacuum can 共125 cm long, 5 cm diameter兲 which can be lowered into a commercial liquid He Dewar. With this arrangement including a few suitably placed copper radiation baffles at the neck of the outer vacuum can, a standard 100 l Dewar lasted for almost a month while heat capacity experiments were carried out in the 4.2– 75 K range. IV. RESULTS AND DISCUSSION

Our observations indicate that due to the high degree of cylindrical symmetry, even though the calorimeter is rather unusually long (⬃10 cm兲, thermometers record the expected temperature values to very high accuracy. Figure 3 shows typical temperature excursions when the calorimeter is filled with pure N2 as a sample. For simplicity we chose square and triangular voltage pulses for two algebraically independent P(t) sequences. Both pulses can be easily generated with the help of a standard signal generator or data acquisition card and a suitable operational amplifier. Before the heat pulse is applied, the calorimeter is left to equilibrate with the surroundings until its temperature trace is horizontal with the time axis. Typically, when the reservoir temperature is adjusted to a new value, the calorimeter attained equilibrium in 30–60 min. To ensure that the calorimeter always attained equilibrium from either a higher temperature or a lower tem⫺ perature for all pulse sequences 共i.e., T→T ⫹ 0 or T→T 0 ), and that small hysteresis of the thermometer does not affect the trace, when T 0 is set to a higher value, a small heat pulse is simultaneously applied to the calorimeter so that its tem-

perature rises above the equilibrium temperature by at least 2–3 K. When such a precaution was not taken, we observed that the base line of the two traces in Fig. 3 differed by ⬃100 mK resulting in spurious results when fit to Eq. 共8兲. Although we assumed that the last term in Eq. 共8兲 will be negligible for ␭ s Ⰷ␭ r , it is clear that for the long exponential decay section of the traces 共below the lower horizontal line in Fig. 3兲 where (T˙ h ⫺T˙ c ) can be very large, the above approximation may not be valid. We observed that for a typical 6.5 K pulse 共similar to Fig. 3兲, the useful window of temperature where the above approximation is valid for obtaining faithful results is only 3.5–4 K 共the section between the two dashed lines in Fig. 3兲. It is crucial that during the entire pulse sequence, the reservoir temperature T 0 should be constant. With the help of a program written in LABVIEW,21 which controls the voltage across the reservoir heater through a multifunction analog-digital/digital-analog card21 and with the thermometer output in the negative feedback loop, we were able to maintain T 0 to within 30 mK during the entire pulse sequence 共see Fig. 3兲. Because the feedback is controlled through software, excellent stability in T 0 was achieved by changing the feedback parameters depending on the pulse height and temperature range. Figure 4 shows the heat capacity curves of the calorimeter filled with pure N2 for two T 0 values. Excellent overlap of the curves demonstrates that with this technique, we can obtain heat capacity of samples with poor thermal conductivity very rapidly (⬃3.5 K range in 3 h兲, without further curve fitting generally required with other techniques.9 By extending the LABVIEW program written for controlling the reservoir temperature T 0 , we fully automated the data acquisition process to obtain pulse sequence data for various T 0 values with an interval of 2.5 K to ensure data overlap from adjacent pulse sequences. Once the data are obtained, standard third degree polynomial fits 共only for the useful temperature interval similar to the one between dashed lines in

Downloaded 19 Dec 2001 to 132.239.1.232. Redistribution subject to AIP license or copyright, see http://ojps.aip.org/rsio/rsicr.jsp

Rev. Sci. Instrum., Vol. 71, No. 10, October 2000

Dual slope heat capacity

3845

to the almost linear temperature dependence of the heat capacity of STYCAST 2850FT near 60 K,19 the presence of the above kink cannot be attributed to the STYCAST O rings used in the construction of the calorimeter. The data shown without further curve fitting clearly demonstrates the high sensitivity of the technique. ACKNOWLEDGMENTS

The authors gratefully acknowledge the support of L. Phelps, G. Labbe, B. Lothrop, W. Malphurs, M. Link, E. Storch, T. Melton, S. Griffin, and R. Fowler. This work is supported by a grant from the National Science Foundation No. DMR-962356. A. Eucken, Z. Phys. 10, 586 共1909兲. W. Nernst, Sitzungsber. K. Preuss. Akad. Wiss. 12, 261 共1910兲. 3 M. T. Alkhafaji and A. D. Migone, Phys. Rev. B 53, 11 152 共1996兲. 4 M. I. Bagatskii, I. Y. Minchina, and V. G. Manzhelii, Sov. J. Low Temp. Phys. 10, 542 共1984兲. 5 L. G. Ward, A. M. Saleh, and D. G. Haase, Phys. Rev. B 27, 1832 共1983兲. 6 P. F. Sullivan and G. Seidel, Phys. Rev. 173, 679 共1968兲. 7 R. Bachmann et al., Rev. Sci. Instrum. 43, 205 共1972兲. 8 S. Riegel and G. Weber, J. Phys. E 19, 790 共1986兲. 9 J. -C. Xu, C. H. Watson, and R. G. Goodrich, Rev. Sci. Instrum. 61, 814 共1990兲. 10 V. K. Pecharsky, J. O. Moorman, and K. A. Gschneidner, Rev. Sci. Instrum. 68, 4196 共1997兲. 11 H. Yao, K. Ema, and C. W. Garland, Rev. Sci. Instrum. 69, 172 共1998兲. 12 J. S. Hwang, K. J. Lin, and C. Tien, Rev. Sci. Instrum. 68, 94 共1997兲. 13 A. J. Jin, C. P. Mudd, N. L. Gershfeld, and K. Fukada, Rev. Sci. Instrum. 69, 2439 共1998兲. 14 E. M. Forgan and S. Nedjat, Rev. Sci. Instrum. 51, 411 共1980兲. 15 R. E. Schwall, R. E. Howard, and G. R. Stewart, Rev. Sci. Instrum. 46, 1054 共1975兲. 16 M. I. Bagatskii, I. Ya. Minchina, and G. Manghelii, Sov. J. Low Temp. Phys. 10, 542 共1984兲. 17 V. G. Manzhelii and Y. A. Freiman, Physics of Cryo Crystals 共AIP, Woodbury, NY, 1997兲. 18 S. Pilla, J. A. Hamida, and N. S. Sullivan, Rev. Sci. Instrum. 70, 4055 共1999兲. 19 C. A. Swenson, Rev. Sci. Instrum. 68, 1312 共1997兲. 20 Scientific Instruments Inc., 4400 W. Tiffany Drive, Mangonia Park, West Palm Beach, FL 33407. 21 National Instruments Corporation, 11500 N. Mopac Expwy., Austin, TX 78759-3504. 1 2

FIG. 5. Heat capacity of the empty cell with small amounts of He exchange gas for better thermal contact. No curve fitting techniques are utilized as the adjacent heat pulse data overlap very well. The sharp kink at 60 K, whose origin is not known, clearly demonstrates the high resolution of the technique near continuous phase transitions.

Fig. 3兲 are employed for obtaining time derivatives in Eq. 共8兲. Once the heat capacity is calculated for each pulse sequence, no further curve fitting is necessary and in general for T⬍30 K, the overlap between adjacent curves is better than the one shown in Fig. 4. Figure 5 shows the heat capacity of the empty calorimeter 共with small amounts of He gas for better thermal contact兲 obtained through the above automated process in the 10– 70 K range. The kink at 60 K, the presence of which is accidental, shows the heat capacity near a continuous phase transition with excellent resolution. Due

Downloaded 19 Dec 2001 to 132.239.1.232. Redistribution subject to AIP license or copyright, see http://ojps.aip.org/rsio/rsicr.jsp

Using RSI format

bath at T0 is represented by the thermal conductance λr . λr may be con- trolled by .... for better thermal contact) obtained through the above auto- mated process in ... 21 National Instruments Corporation, 11500 N. Mopac Expwy., Austin, TX.

82KB Sizes 2 Downloads 255 Views

Recommend Documents

Using RSI format
of the dielectric constant at temperatures ranging from 4.2 to 80 K and in the audio frequency range ... Low frequency dielectric measurements are very impor- .... function generator (Stanford Research Systems Model DS360), (b) isolation.

Using RSI format
1A–1C. A small vessel 250 mL, fitted with a one- way valve and nitrogen gas inlet and liquid nitrogen outlets, is located towards the bottom of the supply Dewar.

Using RSI format
a)Electronic mail: [email protected]. REVIEW .... length providing a 1 mm cylindrical shell space for condens- ... This shell space is vacuum sealed at both.

Using JCP format
Jan 1, 2001 - nature materials are often found confined within host mate- rials. Molecules and ions ... guest molecules insignificant. Table I lists the potential ...

Using JCP format
11–13 Е while the window diameter varies between 4.5 to ... ability per window or bottleneck of the unit cell divided by the probability to be within that cell.

Using PHF format
The non-axisymmetric DNS shows that the streamwise vortex layer developed in the early stage of evolution due to ... and absolute/convective instability have been extensively in- vestigated see Refs. 4–6 and references ... compared with swirl-free

Using JPF1 format
Electron paramagnetic resonance and magnetization in Co doped ... of electron paramagnetic resonance EPR and dc magnetic susceptibility in .... Project No.

Using JPF1 format
In this scenario, revisited early studies of the 1950s4 on mixed-valence manganese perovskites led to the ... VOLUME 89, NUMBER 3. 1 FEBRUARY 2001. 1746.

Using JCP format
degree of micellar organization is characterized by. P(nmax)P(nmin), where nmin and ... are attractive. We also associate an energy b. 0 to each bent bond in the ...

Using JCP format
L. H. Andersen,a) R. Bilodeau,b) M. J. Jensen, S. B. Nielsen, C. P. Safvan,c) and K. Seiersen .... Downloaded 09 Feb 2001 to www.naomib.ca. Redistribution ...

Using JASA format
given degree of confidence that the pulse train has ended. As ... a 2.5-ms separation when Fr 100 Hz.) This condition con- ...... ''Dual temporal pitch percepts.

Using JCP format
Understanding diffusion processes in porous media is a subject of fundamental as ... importance.1–3 Examples of such processes in nature are ion diffusion in ...

Using JEI format
all the binary combinations of these five common carbonates: EC, PC, DMC, EMC, DEC. ... binary carbonate systems: propylene carbonate (PC)-dimethyl car-.

Using JPF1 format
Institute of Physics. S0021-8979(00)09811-X. I. INTRODUCTION ... ing polarization fatigue, becomes critical as the storage ca- pacitor in NV-FRAM decreases in ...

Using JPF1 format
change from single domain to multidomain magnetic behav- ior as grain size increases.16,17 On the other hand, TC remains constant for all the samples see Fig ...

Using JCP format - Henry Fu
The soft-mode frequencies reveal strong bend-stretch coupling in the complex. Excellent .... potential.11 The analysis of the observed transitions is readily.

Using JPF1 format
Department of Physics and Department of Materials Science and Engineering,. North Carolina State University, Raleigh, North Carolina 27695-8202. (Received 5 .... stepped through computer control, and the lock-in outputs are recorded by ...

Using JOE format
E-mail: [email protected]. Abstract. .... the photoexcited carriers in the bulk of the material after they ... in heat generation, followed by bulk and surface recombi-.

Using JPF1 format
May 1, 2000 - host atoms by the spin-orbit interaction, which allows the impurity spin to get ... of the conduction electrons on the nonmagnetic host atoms.

Using JCP format
Run III attempts to simulate the β- to α-phase transformation in the NPT ensemble while run IV is an NPT ensemble simulation of liquid toluene. Run IIIa is an.

Using JPF1 format
Feb 1, 2001 - Magnetotransport properties of manganite/insulator composites are studied in this article. By merging the half metallic character of ...

The effectiveness of rapid sequence intubation (RSI) versus non-RSI ...
Page 1 of 9. ORIGINA L R E S EA RCH Open Access. The effectiveness of rapid sequence. intubation (RSI) versus non-RSI in. emergency department: an analysis of. multicenter prospective observational study. Masashi Okubo1*, Koichiro Gibo2. , Yusuke Hag

RSI 087-2015.pdf
Page 1 of 14. Page 2 of 14. Page 2 of 14. Page 3 of 14. Page 3 of 14. RSI 087-2015.pdf. RSI 087-2015.pdf. Open. Extract. Open with. Sign In. Main menu.

RSI 130-2014.pdf
Page 1 of 12. Page 1 of 12. Page 2 of 12. Page 2 of 12. Page 3 of 12. Page 3 of 12. RSI 130-2014.pdf. RSI 130-2014.pdf. Open. Extract. Open with. Sign In.Missing: