First Sensor PIN PD Data Sheet Part Description X100-7 THD International / US Order # 501400 / 01-166 Version 12-11-13
Features
Description
Application
RoHS
• 100 mm² PIN detector • Light blocking encapsulant • Low dark current • Low capacitance • High shunt resistance • High sensitivity
Square active area PIN photodiode with 100 mm² active area. Ceramic carrier type 2-pin package with light blocking black epoxy encapsulant. Reflow solderable. Non-hermetic.
• Ionizing radiation detector • Medical equipment • Personal dosimeter
2002/95/EC
Absolute maximum ratings Parameter Storage temp Operating temp Max reverse voltage Peak DC current
Absorption of gamma radiation (23 °C) Min -40 -20
Max 100 70 50 10
Unit °C °C V mA
100,0%
Absorption probability (%)
Symbol TSTG TOP Vmax IPEAK
Schematic Pin 2
10,0%
1,0%
0,1% 1
Pin 1
10
100
1000
Gamma energy (keV)
Electro-optical characteristics @ 23 °C Symbol
ID TK(ID) C tR
VBR
Characteristic Active area Active area Energy range of detectable radiation Dark current Temperature coefficient Capacitance Rise time Shunt Resistance Noise current Breakdown voltage
European, International Sales:
Test Condition
Typ 10 x 10 100
5
Gamma radiation VR = 10 V VR = 10 V; change of dark current VR = 0 V; f = 10 kHz VR = 10 V; f = 10 kHz VR = 12 V; E = 10 keV; RL = 50 Ω VR = 10 mV VR = 12 V IR = 2 µA
First Sensor AG Peter-Behrens-Strasse 15 12459 Berlin Germany T +49 30 6399 2399 F +49 30 639923-752
[email protected]
Min
USA:
Max
1000 1.5 13 500 90
3
500
50
40 6.1 E-14 80
First Sensor Inc. 5700 Corsa Avenue #105 Westlake Village CA 91362 USA T +1 818 706 3400 F +1 818 889 7053
[email protected]
Unit mm mm² keV nA %/K pF pF ns MΩ A/√Hz V
First Sensor PIN PD Data Sheet Part Description X100-7 THD International / US Order # 501400 / 01-166 Version 12-11-13
Dark current (23 °C)
Capacitance as fct of reverse bias (23 °C)
1,0E-07
500 450
1,0E-08
Capacitance (pF)
Dark current (A)
400
1,0E-09
350 300 250 200 150 100 50
1,0E-10
0 0
10
20
30
0
10
20
30
40
Reverse bias (V)
Reverse bias (V)
Package dimension: Small quantities: Foam pad, boxed (12 cm x 16.5 cm)
Handling: Please refer to document "Instructions for handling and processing"
Disclaimer:
Due to our strive for continuous improvement, specifications are subject to change within our PCN policy according to JESD46C.
European, International Sales:
First Sensor AG Peter-Behrens-Strasse 15 12459 Berlin Germany T +49 30 6399 2399 F +49 30 639923-752
[email protected]
USA:
First Sensor Inc. 5700 Corsa Avenue #105 Westlake Village CA 91362 USA T +1 818 706 3400 F +1 818 889 7053
[email protected]