First Sensor Ionizing Radiation Detector Data Sheet Part Description X5-ɣ TO Order # 50190301 Version 14-09-12
Foto
Features
Description
Application
RoHS
• 5 mm² PIN detector • Very low dark current • Very low dark capacitance • Fully depletable • Guard ring diode
Circular active area PIN photodiode with 5 mm² active area. Metal can type non-hermetic TO8Si package with aluminzed Kapton window.
• Dosimeter • Radiation monitoring • Gamma diffraction
2002/95/EC
Symbol TSTG TOP UR IPEAK
Parameter Storage temp Operating temp Reverse voltage Peak DC current
Gamma - Absorption (23 °C) Min -55 -40
Max 125 100 300 10
Unit °C °C V mA
Schematic
100% 90% 80%
Gamma Absorption
Absolute maximum ratings
70% 60% 50% 40% 30% 20% 10% 0% 0
5
10
15
20
25
30
35
40
45
50
Gamma energy (keV)
Electro-optical characteristics @ 23 °C Symbol
ID
Characteristic No of elements Active area Window material Scintillator Dark current
C
Capacitance
tR UBR
Rise time Breakdown voltage Energy range of detectable radiation
European, International Sales:
Test Condition Diameter
U=0V U = 120 V (full depletion) U=0V U = 120 V (full depletion) U = 120 V; 20 keV; RL = 50 Ω IR = 2 µA Alpha radiation Beta radiation X-Ray / Gamma radiation First Sensor AG Peter-Behrens-Strasse 15 12459 Berlin Germany T +49 30 6399 2399 F +49 30 639923-752
[email protected]
USA:
Min
Typ Max Unit 1 2.52 mm double-aluminized Kapton (8 µm) none <1 pA 10 pA 35 pF 2.5 pF 15 ns 300 500 V 2 MeV 20 1000 keV 1 50 keV First Sensor Inc. 5700 Corsa Avenue #105 Westlake Village CA 91362 USA T +1 818 706 3400 F +1 818 889 7053
[email protected]
First Sensor Ionizing Radiation Detector Data Sheet Part Description X5-ɣ TO Order # 50190301 Version 14-09-12
Dark current as fct of reverse bias (23 °C)
Capacitance as fct of reverse bias (23 °C)
1,E-10
30,0
1,E-11
Capacitance (pF)
Dark current (A)
25,0
1,E-12
1,E-13
20,0 15,0 10,0 5,0 0,0
0
10
20
30
40
50
60
70
80
90
100 110 120
0
Reverse bias (V)
10
20
30
40
50
60
70
80
90
100 110 120
Reverse bias (V)
Package dimension Small quantities:Foam pad, boxed (12 cm x 16.5 cm)
Handling:
Please refer to document "Instructions for handling and processing", Please consider basic ESD protection while handling.
Disclaimer:
Due to our strive for continuous improvement, specifications are subject to change within our PCN policy according to JESD46C.
European, International Sales:
First Sensor AG Peter-Behrens-Strasse 15 12459 Berlin Germany T +49 30 6399 2399 F +49 30 639923-752
[email protected]
USA:
First Sensor Inc. 5700 Corsa Avenue #105 Westlake Village CA 91362 USA T +1 818 706 3400 F +1 818 889 7053
[email protected]