First Sensor Ionizing Radiation Detector Data Sheet Part Description X5-ɣ TO Order # 50190301 Version 14-09-12

Foto

Features

Description

Application

RoHS

• 5 mm² PIN detector • Very low dark current • Very low dark capacitance • Fully depletable • Guard ring diode

Circular active area PIN photodiode with 5 mm² active area. Metal can type non-hermetic TO8Si package with aluminzed Kapton window.

• Dosimeter • Radiation monitoring • Gamma diffraction

2002/95/EC

Symbol TSTG TOP UR IPEAK

Parameter Storage temp Operating temp Reverse voltage Peak DC current

Gamma - Absorption (23 °C) Min -55 -40

Max 125 100 300 10

Unit °C °C V mA

Schematic

100% 90% 80%

Gamma Absorption

Absolute maximum ratings

70% 60% 50% 40% 30% 20% 10% 0% 0

5

10

15

20

25

30

35

40

45

50

Gamma energy (keV)

Electro-optical characteristics @ 23 °C Symbol

ID

Characteristic No of elements Active area Window material Scintillator Dark current

C

Capacitance

tR UBR

Rise time Breakdown voltage Energy range of detectable radiation

European, International Sales:

Test Condition Diameter

U=0V U = 120 V (full depletion) U=0V U = 120 V (full depletion) U = 120 V; 20 keV; RL = 50 Ω IR = 2 µA Alpha radiation Beta radiation X-Ray / Gamma radiation First Sensor AG Peter-Behrens-Strasse 15 12459 Berlin Germany T +49 30 6399 2399 F +49 30 639923-752 [email protected]

USA:

Min

Typ Max Unit 1 2.52 mm double-aluminized Kapton (8 µm) none <1 pA 10 pA 35 pF 2.5 pF 15 ns 300 500 V 2 MeV 20 1000 keV 1 50 keV First Sensor Inc. 5700 Corsa Avenue #105 Westlake Village CA 91362 USA T +1 818 706 3400 F +1 818 889 7053 [email protected]

First Sensor Ionizing Radiation Detector Data Sheet Part Description X5-ɣ TO Order # 50190301 Version 14-09-12

Dark current as fct of reverse bias (23 °C)

Capacitance as fct of reverse bias (23 °C)

1,E-10

30,0

1,E-11

Capacitance (pF)

Dark current (A)

25,0

1,E-12

1,E-13

20,0 15,0 10,0 5,0 0,0

0

10

20

30

40

50

60

70

80

90

100 110 120

0

Reverse bias (V)

10

20

30

40

50

60

70

80

90

100 110 120

Reverse bias (V)

Package dimension Small quantities:Foam pad, boxed (12 cm x 16.5 cm)

Handling:

Please refer to document "Instructions for handling and processing", Please consider basic ESD protection while handling.

Disclaimer:

Due to our strive for continuous improvement, specifications are subject to change within our PCN policy according to JESD46C.

European, International Sales:

First Sensor AG Peter-Behrens-Strasse 15 12459 Berlin Germany T +49 30 6399 2399 F +49 30 639923-752 [email protected]

USA:

First Sensor Inc. 5700 Corsa Avenue #105 Westlake Village CA 91362 USA T +1 818 706 3400 F +1 818 889 7053 [email protected]

x5-y-50190301.pdf

2 μA 300 500. tR. 5700 Corsa Avenue #105. Westlake Village. CA 91362 USA. T +1 818 706 3400. F +1 818 889 7053. First Sensor AG. Peter-Behrens-Strasse 15. 12459 Berlin. Germany. T +49 30 6399 2399. F +49 30 639923-752. Alpha radiation. X-Ray / Gamma radiation 1 50. 2. First Sensor Inc. Beta radiation 20 1000.